• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON & CO (FRANKLIN LAKES) BD VACUTAINER® SST¿ II ADVANCE; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON & CO (FRANKLIN LAKES) BD VACUTAINER® SST¿ II ADVANCE; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION Back to Search Results
Catalog Number 367953
Medical Device Problem Code Device Ingredient or Reagent Problem (2910)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/17/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported while using bd vacutainer® sst¿ ii advance, an unspecified number of tubes contained oil gel globules and gel smearing after processing that clogged their instrument probe.There was no health impact or consequences reported.
 
Event or Problem Description
It was reported while using bd vacutainer® sst¿ ii advance, an unspecified number of tubes contained oil gel globules and gel smearing after processing that clogged their instrument probe.There was no health impact or consequences reported.
 
Additional Manufacturer Narrative
Investigation summary bd received 1 video for investigation which indicated gel smearing; oil gel globules were not observed.100 retained samples were visually inspected, and no gel defects were found.Complaints for sample quality were not under statistical control for the month of december 2025, additional testing will be performed.Factors that may contribute to gel smearing and oil gel globule were evaluated through a clinical study to verify the design of the device met it¿s intended use.Bd was unable to duplicate the customer's indicated failure modes, gel smearing and oil gel globule, via clinical investigation because the defects were not evident in the testing of the complaint lot samples.All visual observations of both retain and control samples demonstrated clinically acceptable performance.The result of the study showed that the device performed as expected and we were unable to determine any external contributor to this reported issue.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure modes gel smearing vial video analysis.This complaint has not been confirmed for the indicated failure mode oil gel globule.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for the identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SST¿ II ADVANCE
Common Device Name
TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION
Manufacturer (Section D)
BECTON DICKINSON & CO (FRANKLIN LAKES)
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
anand singh
1 becton drive
franklin lakes, NJ 07417
2018475186
MDR Report Key24125931
Report Number9617032-2025-02460
Device Sequence Number15795650
Product Code JKA
UDI-Device Identifier30382903679530
UDI-Public(01)30382903679530(17)261031(10)5121194
Combination Product (Y/N)N
Initial Reporter CountryPE
PMA/510(K) Number
BK050036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number367953
Device Lot Number5121194
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2025
Supplement Date Received by Manufacturer03/20/2026
Initial Report FDA Received Date01/20/2026
Supplement Report FDA Received Date04/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/01/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-