| Model Number |
MR850JHU |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The subject mr850 respiratory humidifier has been requested to be returned to fisher & paykel healthcare new zealand for evaluation.We will provide a follow up report upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in (b)(6) reported that the audible alarm of an mr850 respiratory humidifier was not functioning.There was no patient involvement reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(6).Corrections: section d1: brand name has been updated.Section h4: device manufacture date corrected.Section h11: method: the subject pcb of the mr850 respiratory humidifier was returned to fisher & paykel (f&p) healthcare in new zealand for investigation where it was performance tested.Our investigation is based on the evaluation of the subject pcb, information provided by the customer and our knowledge of the product.Results: testing of the subject pcb confirmed that no sound was generated from the speaker.Additional resistance testing identified an open circuit in the speaker's coil.Conclusion: the cause of the speaker failure was an open circuit in the speaker coil.The mr850 respiratory humidifier product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking and performance testing of the mr850 respiratory humidifier.In addition, the product technical manual also states that "all servicing procedures should be followed by a humidifier test, and an electrical safety test to ensure proper operation".The mr850 respiratory humidifier is equipped with visual alarm indicators in addition to the audible alarm.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in illinois reported that the audible alarm of an mr850 respiratory humidifier was not functioning.There was no patient involvement reported.
|
| |
|