|
This supplemental report is being submitted to provide the results of the final investigation.Updated field(s): d4, d8, d9, h2, h3, h4, h6, h11.Correction h6 from a02 to a041001.Correction h6 from g04092 to g04023.The device was returned to olympus for inspection, and the reported failure was confirmed.Based on the results of the investigation, olympus confirmed the cause was due to the tip of the insertion part, the coil sheath was deformed and there was a hole in the outer tube, however the most probable cause of the complaint is cause not established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|