ALCON RESEARCH, LLC - HUNTINGTON CLAREON ASPHERIC UV ABSORBING IOL; INTRAOCULAR LENS Back to Search Results
Model Number CC60WF
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/02/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product was not returned for analysis.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-healthcare professional reported during intraocular lens (iol) implant procedure, upon implantation, surgeon noticed that the lens was cracked and had to explant lens and insert a replacement during same procedure.Additional information has been requested and received stating in the surgeon¿s opinion, product caused or contributed to the product malfunction.There was no patient harm.
 
Additional Manufacturer Narrative
Additional information was provided in d.9., h.3., h.6.And h.11.The lens was returned in two pieces inside the lens case.One portion was returned in the well area.One portion was returned on top of a post of the lens case.This caused further damage to the lens.Viscoelastic was observed on the lens.The optic had been cut into two pieces across the center.The lens was cleaned with klrp for further evaluation.The portion damage by the lens case post had other cracked areas.This damage was similar in appearance to damage that can occur if there is a plunger override.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.Associated products were not provided.It is unknown if qualified products were used.The root cause for cracked optic cannot be determined.The lens was returned in two pieces inside the lens case.One portion was returned on top of a post of the lens case.This caused further damage to the lens.The portion damage by the lens case post had other cracked areas.This damage was similar in appearance to damage that can occur if there is a plunger override.It is unknown if qualified associated products were used.The instructions for use (ifu) instructs: company foldable intraocular lenses (iols) are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly, causing delivery issues and/or damage.Important: the plunger should make initial contact with the cartridge at the ramp.In the event the plunger does not contact the cartridge at the ramp, do not use the handpiece and contact company.The handpiece ifu instructs: verify that the plunger tip is properly engaging the lens optic and that the lens moves forward at the same rate as the plunger while slowly advancing the plunger forward to avoid damaging the lens.When the threads on the knob make contact with the barrel, turn the knob clockwise approximately half turn to engage the threads and then stop.The iol will now be in the dwell position.Inspect to ensure the plunger is behind the optic.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC UV ABSORBING IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24129239
Report Number1119421-2026-00158
Device Sequence Number20197667
Product Code HQL
UDI-Device Identifier00380652396875
UDI-Public(01)00380652396875(17)290618(11)250620(10)16059279
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCC60WF
Device Lot Number16059279
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/29/2025
Supplement Date Received by Manufacturer02/02/2026
Initial Report FDA Received Date01/21/2026
Supplement Report FDA Received Date02/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/20/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN CARTRIDGE.; UNKNOWN INJECTOR/HANDPIECE.; UNKNOWN LOADING/FOLDING FORCEPS.; UNKNOWN VISCOELASTIC.
Patient SexUnknown
-
-