Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the tube presented a kink near the 18cm marking, besides the kink the tube presented a damage.A dimensional test was performed inner diameter was measured 0.279 inch this reading is within specification (0.276 +/- 0.008 inch), and outer diameter was measured 0.376 this reading is within specification (0.374 +/- 0.008 inch).It was reported that during use, the endotracheal tube was twisted and bent at the patient's larynx, obstructing airflow and making suctioning impossible.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: avoid exposure to elevated temperatures and ultraviolet light during storage.Test the cuff, pilot balloon and valve for integrity by inflation prior to use.Insert a luer tip syringe into the cuff inflation valve housing and inject enough air to fully inflate cuff.After test inflation, completely evacuate the air.If shortening of the endotracheal tube by cutting is considered, the tube should be cut and the connector reinserted prior to intubation.Tubes with 15mm connectors that cannot be removed with reasonable manipulation are not suitable for cutting.Always assure the connector is firmly seated in both the tracheal tube and the breathing circuit to prevent disconnection during use.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|