| Model Number |
MI1250 SYNCHRONY 2 |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue wire breakages.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|
Event or Problem Description
|
|
The device was explanted due to poor sound quality.The user has been reimplanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been explanted and has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
| |
|
Event or Problem Description
|
|
The device was explanted due to poor sound quality.The user has bene reimplanted.
|
| |
|