• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INCEPTA? ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INCEPTA? ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) Back to Search Results
Model Number E163
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) system exhibited shock impedance measurements out of range on the right ventricular (rv) lead.It was noted that the right ventricular (rv) lead was not a boston scientific product.Technical services (ts) provided assistance.This device remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INCEPTA? ICD
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key24130084
Report Number2124215-2026-03804
Device Sequence Number20176099
Product Code LWP
UDI-Device Identifier00802526480782
UDI-Public(01)00802526480782(17)161205
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Remedial Action Recall,Notification
Type of Report Initial,Followup
Report Date (Section B) 06/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/05/2016
Device Model NumberE163
Device Catalogue NumberE163
Device Lot Number110051
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2025
Supplement Date Received by Manufacturer06/10/2026
Initial Report FDA Received Date01/21/2026
Supplement Report FDA Received Date06/15/2026
Date Device Manufactured10/23/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age38 YR
Patient SexFemale
-
-