| Model Number |
EVPROPLUS-23US |
| Medical Device Problem Codes |
Calcified (1077); Gradient Increase (1270); Thickening of Material (4056); Device Stenosis (4066)
|
| Health Effect - Clinical Codes |
Valvular Stenosis (2697); Mitral Valve Insufficiency/ Regurgitation (4451); Tricuspid Valve Insufficiency/ Regurgitation (4453); Transvalvular Pressure Gradient Increased (4854)
|
| Date of Event |
10/24/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that approximately three years, nine and a half months following the implant of a medtronic transcatheter aortic bioprosthetic valve (tav), a multi-detector computed tomography was performed.Structural valve deterioration was reported; predominant aortic stenosis was the primary mode of valve failure.This was characterized by an increase in mean gradient of >=20 mmhg from baseline and a mean gradient of >=30 mmhg.The patient was enrolled in a clinical study with existing, baseline valve failure to determine whether a redo-tav replacement procedure, tav-in-tav, was appropriate.Approximately two and a half months later, reintervention was performed.A non-medtronic valve was implanted tav-in-tav.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated data: b5, h6 corrected data: a2, h6 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated that prior to the valve tav-in-tav revision, the existing transcatheter valve did have leaflet thickening, but a thrombus was not present.
|
| |
|
Search Alerts/Recalls
|