| Model Number |
VTICMO 13.2 (-14.50/ 1.0) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
12/12/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced a refractive surprise.In a separate surgery, the lens was exchanged with the different model/ same length lens and the problem was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
H6- investigation type code: 4110- lens work order search-no similar complaint event(s) within associated lots were found.H11- manufacturer narrative: secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5.- a facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced over/under correction.In a separate surgery the lens was exchanged with the different model/ same length, weaker power lens and the problem was resolved h6.Health effect- clinical code (e): '2137' should be removed.'4581- over/under correction' should be added.Claim# (b)(4).
|
| |
|