| Model Number |
DDU-100A |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported their aed is flashing red and prompting a "replace battery pack now" warning.They reported this did not occur during patient use.
|
| |
|
Additional Manufacturer Narrative
|
|
Troubleshooting of the aed with the customer identified that once the new battery was installed and a manual self test run the aed was functioning as designed and could stay in service.Although requested, the log files from the device have not been provided and the cause of the complaint is not known.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the log files from the aed identified that the customer was performing excessive self-testing of the aed which reduced the battery life expectancy.As a follow-up with the customer, the proper maintenance of this device was reviewed.
|
| |
|