| Catalog Number |
367989 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There were multiple 510k numbers reported to be involved.The information is as follows: g.4.Pma / 510(k)#: k230855.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® sst¿, 1 tube was cracked prior to processing.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received 2 photos for investigation which show a tube with a crack along its side.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode damaged tube.No definitive root cause could be established for the reported defect.Based on the low defect rate for the batch in question, no actions are planned at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® sst¿, 1 tube was cracked prior to processing.There was no health impact or consequences reported.
|
| |
|