| Model Number |
P005723CAP02 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The baxter technician found the power cord needed to be replaced.It is necessary for the p500 to have an effective maintenance program.We recommend that you do annual preventive maintenance (pm) and testing for joint commission certification.Pm and testing not only meet joint commission requirements but can help make sure of a long, operative life for the p500.Pm will minimize downtime due to excessive wear.Examine the plug for damage.Make sure the plug is a one-piece molded plug assembly.If it is not, replace the plug cord assembly.Replace any plug cord assembly that shows any of these: discoloration of the plug molding around the plug blades; this could occur if the plug blades have overheated or arced.Any signs of cracking; this could occur if the plug has been bent and straightened to a point past its useful life.Loose fit of the plug blade (the plug blade moves in the molding); this could occur if the molding has overheated or the blades have been bent and straightened to a point past their useful life.Replace the power cord, if damaged.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the power cord to resolve the reported event.Based on this information, no further action is required.
|
| |
|
Event or Problem Description
|
|
On 29-dec-2025, a company representative - service personnel contacted technical service to report that p500 mrs control unit (product code p005723cap02, serial number (b)(6)), had power cord damaged with copper wire exposed.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
| |
|
Search Alerts/Recalls
|