| Catalog Number |
221728051 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that saw blade made contact with trial, putting a slice in it.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional manufacturer narrative: e1: please note that the account name and address were not provided.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d4 (lot ),d9 ,h4.Corrected: d4 (primary udi).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: according to the information received, ¿it was reported that saw blade made contact with trial, putting a slice in it¿.The product returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis revealed that pinnacle mb head trial 51_28 exhibited a vertical deep cut on the surface.No additional defects were identified.The damage observed in the provided evidence is consistent with a saw blade making an untended contact with the device, a condition that corresponds with the information previously reported.Properly handling and attention to the approved use of the device diminishes the risk of failure.A dimensional inspection and functional testing were not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the pinnacle mb head trial 51_28 would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Corrected: h3.
|
| |
|
Search Alerts/Recalls
|