ALCON RESEARCH, LLC - HOUSTON VISCOUS FLUID CONTROL TUBING SET (SILICONE OIL INJ/EXT); TUBING, NONINVASIVE Back to Search Results
Catalog Number 8065750957
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported that the tubing leaked occurred during vitrectomy surgery.The surgery was completed on the same day after replacing the product with another one.There was no impact to the patient.
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.The unused viscous fluid control (vfc) tubing set, small parts tray (smpt) and 10millilitre (ml) syringe were returned and visually inspected.In return condition, the gray tip plunger was inside the smpt.The vfc tubing set, 10ml syringe, and smpt were found to be in good condition.The 25 guage (ga), 23ga and 20ga vfc cannula needle tips could fit into 10ml syringe.The vfc tubing set and gray tip plunger could fit into lab 10ml syringe.The vfc tubing set was tested using a calibrated console.The light emitting diode rings on the console turned green as the gray connector was engaged to the console indicating the proper communication between the connector and the console.The plunger performed smoothly; radio frequency identification (rfid) connection to the console and the syringe to the adaptor/tubing securely engaged, not detached.The pressure was stable in both injection and extraction settings.No air leak occurred from the sample during testing.The investigation conducted a non-conformance review of the reported batch number.No deviations that could have contributed to this event were identified and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.The product functioned per specifications.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.No further investigative or manufacturing actions are warranted at this time due to the product meeting specifications.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISCOUS FLUID CONTROL TUBING SET (SILICONE OIL INJ/EXT)
Common Device Name
TUBING, NONINVASIVE
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24141768
Report Number1644019-2026-00476
Device Sequence Number21207797
Product Code GAZ
UDI-Device Identifier00380657509577
UDI-Public(01)00380657509577(17)270731(11)250815(10)17KN5C
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065750957
Device Lot Number17KN5C
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/30/2025
Supplement Date Received by Manufacturer04/30/2026
Initial Report FDA Received Date01/21/2026
Supplement Report FDA Received Date05/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/15/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-