| Catalog Number |
VC10A256F100 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
11/06/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent a radio frequency ablation procedure using the venclose catheter.During the procedure, the account experienced multiple ehits following procedures with newly reintroduced catheters, totaling at least six incidents in the past month.All catheters used were new since market return.Rvt confirmed applying pressure along the catheter¿s heating element during treatment, with two burns noted at the proximal end.Pre-treatment measurements averaged 6.8 cm from the terminal valve at the sfj to the catheter tip.The procedure was completed using another catheter.Current status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported thrombosis could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.D4 (unique identifier (udi) #), g3, h6 (method).Section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent a radio frequency ablation procedure using the venclose catheter.During the procedure, the account experienced multiple ehits following procedures with newly reintroduced catheters, totaling at least six incidents in the past month.All catheters used were new since market return.Rvt confirmed applying pressure along the catheter¿s heating element during treatment, with two burns noted at the proximal end.Pre treatment measurements averaged 6.8 cm from the terminal valve at the sfj to the catheter tip.The procedure was completed using another catheter.The current status of the patient is unknown.
|
| |
|
Search Alerts/Recalls
|