| Model Number |
1420 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that a review of log file data revealed the controller exhibited an unexpected loss of power.The controller remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Event or Problem Description
|
|
It was further reported that due to a patient error a double disconnect occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental report is being submitted for additional event details.Updated b5 and h6.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental report is being submitted for device evaluation.Product event summary: the controller ((b)(6)) was not returned for evaluation.Review of the manufacturing documentation confirmed that the associated device met all requirements for release.Log file analysis revealed a controller power up event with an associated pump start event was logged on (b)(6) 2026 at 18:51:00, indicating a loss of power to the controller.The data point prior to the loss of power revealed that (b)(6) was connected to power port one (1) with (b)(4) relative state of charge (rsoc) and (b)(6) was connected to power port two (2) with (b)(4) rsoc.The data point recorded after the loss of power revealed that (b)(6) was connected to power port one (1) and (b)(6) was connected to power port two (2).The controller was without power for approximately ten (10) seconds.As a result, the reported loss of power event was confirmed.The most likely root cause of the loss of power event can be attributed to the double disconnection of power sources, as described in the event details.Capa pr00551638 investigated controller losses of power.Even though this capa is now closed, (b)(6) falls in scope of this capa.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|