KARL STORZ ENDOVISION VIDEO URETERO-RENOSCOPE FLEX-XC1 Back to Search Results
Model Number 091271-06
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that flex-xc1 091271-06 during rirs procedures suddenly went out.The vigilance case involved 2 pieces of the same device with the same lot number as they were used during the same surgical procedure.There was a delay of procedure of less than 15 minutes.No negative impact in state of health reported.
 
Additional Manufacturer Narrative
Result of investigation: as the endoscopes have not been sent back, a final determination of the root cause is not possible.As the customer describes that both worked for a while, it is most likely that they got damaged during use, by e.G.Application of too much force during the procedure.The above investigations did not reveal a root cause related to the labelling, the device, or its history.All production-related quality checks were passed, and no non-conformities were identified in the device's manufacturing records.The labelling was found to contain adequate instructions and warnings.The complaint history did not reveal any pickup in similar complaints; no trend was identified.The event is filed under internal karl storz complaint id: (b)(4).
 
Additional Manufacturer Narrative
This device report was created in error.Please refer to (b)(4).The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEO URETERO-RENOSCOPE FLEX-XC1
Common Device Name
VIDEO URETERO-RENOSCOPE FLEX-XC1
Manufacturer (Section D)
KARL STORZ ENDOVISION
91 carpenter hill road
charlton MA 01507
Manufacturer (Section G)
KARL STORZ ENDOVISION
91 carpenter hill road
charlton MA 01507
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245
MDR Report Key24148994
Report Number1221826-2026-00081
Device Sequence Number8841179
Product Code FGB
UDI-Device Identifier04048551428009
UDI-Public04048551428009
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K212458
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number091271-06
Device Catalogue Number091271-06
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/07/2026
Supplement Date Received by Manufacturer01/07/2026
Initial Report FDA Received Date01/22/2026
Supplement Report FDA Received Date01/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-