AIZU OLYMPUS CO., LTD. BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-1T150
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/17/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection.Based on the results of the investigation, it¿s likely the loss of image occurred due to an electronic component failure.However, a definitive root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was observed that during the device evaluation, the bronchovideoscope exhibited a loss of image.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRONCHOVIDEOSCOPE
Common Device Name
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key24150238
Report Number9610595-2026-07646
Device Sequence Number9335381
Product Code EOQ
UDI-Device Identifier04953170288968
UDI-Public04953170288968
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K023984
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberBF-1T150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/09/2026
Initial Report FDA Received Date01/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-