This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d9, h2, h3, h6, h11.The device was returned to olympus for inspection, and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: missing screw on coupler, lines showing on image and failed leak test.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: missing screw on coupler, lines showing on image due to faulty charged coupler device and failed leak test due to cut on cable.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|