Catalog Number 115-322
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/10/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Reason for reoperation: rupture.
 
Event or Problem Description
Healthcare professional reported "exchange due to concerns with the product" and "rupture".This record is for the left side.Device remains implanted.
 
Additional Manufacturer Narrative
Additional, corrected and/or changed data: b.5, d.6b, h.6.
 
Event or Problem Description
Healthcare professional reported "exchange due to concerns with the product" and "rupture".This record is for the left side.Device was explanted.
 
Additional Manufacturer Narrative
Device evaluation: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: rupture: observed broken device through microscopic inspection assessed as assessed as fold flaw opening.Observed a missing piece of shell through microscopic inspection assessed as inconclusive.No further actions are required as it is not related to the manufacturing process.Anxiety - product/ procedure: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (wear abrasion and crease) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
 
Event or Problem Description
Healthcare professional reported "exchange due to concerns with the product" and "rupture".This record is for the left side.Device was explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 115 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key24153040
Report Number9617229-2026-01398
Device Sequence Number10058930
Product Code FTR
UDI-Device Identifier10888628002388
UDI-Public10888628002388
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/17/2016
Device Catalogue Number115-322
Device Lot Number2135414
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/06/2026
Supplement Date Received by Manufacturer02/05/2026
04/15/2026
Initial Report FDA Received Date01/22/2026
Supplement Report FDA Received Date02/25/2026
05/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/17/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
Patient SexFemale
Patient Weight91 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-