| Catalog Number |
D134804 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Cardiac Tamponade (2226)
|
| Date of Event |
12/30/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: pc-(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an atrial flutter (afl) procedure with a thermocool smarttouch sf catheter and suffered a pericardial effusion which required a pericardiocentesis.While ablating the cti line in the right isthmus, two steam pops were heard.The patient became hypotensive.Carto sound was used and a pericardial effusion was noted.A pericardiocentesis was performed and approximately 425cc of fluid was extracted.Further intervention is unknown at this time.Hardware worked within specifications.Multiple attempts have been made to obtain clarification of this complaint.However, no further information has been made available.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was completed on 31-mar-2026.An analysis of the product could not be performed since a physical sample was not received for evaluation.A manufacturing record evaluation was performed for the finished device number lot 31662217l and no internal actions related to the complaint were found during the review.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Based on the information available, there is no indication that a design or manufacturing issue has caused the reported complaint condition.If the product is received at a later date, the investigation will be updated as applicable.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|