MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ENDURITY DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ENDURITY DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number PM2160
Medical Device Problem Codes Failure to Interrogate (1332); No Pacing (3268)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/13/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient had presented to the clinic for a routine follow-up.It was noted that the pacemaker had failed to be interrogated via inductive telemetry.There was no lv output, and pacing could not be induced with magnet application.The pacemaker was subsequently explanted and replaced.The patient remained in stable condition.
 
Additional Manufacturer Narrative
The reported events of no lv output, no inductive telemetry, no magnet response were confirmed.The device was received with no telemetry communication and no output.Visual inspection of the header attachment area detected an anomaly between the epoxy header and titanium case.The device was cut open to enable further testing and the battery was found at below eos levels.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Feedthrough leak test was performed, indicating a device hermeticity breach.This is consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to internal electronics.Hybrid circuitry was tested by connecting to an external power source.Test results indicated elevated current drain, consistent with moisture damage, resulting in the reported event.A manufacturing process anomaly consistent with header bonding anomaly may have occurred.The device is included in the assurity and endurity pacemakers' header anomaly advisory issued by abbott on 15 march 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURITY DR
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key24157770
Report Number2017865-2026-01495
Device Sequence Number9113209
Product Code NVZ
UDI-Device Identifier05414734507059
UDI-Public(01)05414734507059(10)A000017281(17)170731
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 04/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2017
Device Model NumberPM2160
Device Lot NumberA000017281
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2026
Supplement Date Received by Manufacturer03/17/2026
Initial Report FDA Received Date01/22/2026
Supplement Report FDA Received Date04/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1531-2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-