• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLACK FOREST MEDICAL GMBH DORO® SKULL CLAMP; DORO® QR3 SKULL CLAMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLACK FOREST MEDICAL GMBH DORO® SKULL CLAMP; DORO® QR3 SKULL CLAMP Back to Search Results
Model Number 1001.001
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code Laceration(s) (1946)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
There is no apparent connection between the deviation detected in this investigation (loose threaded insert on the fixed arm) and the reported slippage.Any deviations due to non-compliance with the maintenance cycle specified in the ifu may remain undetected if inspection and maintenance are not carried out, therefore the reporting customer is advised of the importance of complying with the maintenance interval within this complaint correspondence.From our experience, pinning technique plays a significant role in slippage during skull fixation, so it cannot be excluded that the pinning of the patient's skull may not have been optimal as described in the instruction manual: "adjust the skull clamp to the width of the patient's head in the manner that the two skull pins in the rocker arm are equidistant from the centerline of the head and the single skull pin at the extension assembly is in line with this centerline.".
 
Event or Problem Description
Customer informed us on (b)(6) that one of our products was involved in a case in which the patient sustained a laceration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORO® SKULL CLAMP
Common Device Name
DORO® QR3 SKULL CLAMP
Manufacturer (Section D)
BLACK FOREST MEDICAL GMBH
boetzinger strasse 86
freiburg im breisgau, baden-wuerttemberg 79111
GM  79111
Manufacturer (Section G)
BLACK FOREST MEDICAL GMBH
boetzinger strasse 86
freiburg im breisgau, baden-wuerttemberg 79111
GM   79111
Manufacturer Contact
nicholas preissler
boetzinger strasse 86
freiburg im breisgau, baden-wuerttemberg 79111
GM   79111
MDR Report Key24158741
Report Number3003923584-2026-00002
Device Sequence Number9590046
Product Code HBL
UDI-Device Identifier04250435506196
UDI-Public04250435506196
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K203505
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Repair,Inspection
Type of Report Initial
Report Date (Section B) 01/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1001.001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/26/2025
Initial Report FDA Received Date01/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/25/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-