| Model Number |
0684-00-0479-07 |
| Medical Device Problem Code |
Pressure Problem (3012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Due to characterization limit e1: initial reporter: dr.(b)(6).Event site name: (b)(6) hospital.Event site address: (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that linear 7.5fr.34cc intra-aortic balloon pump (iabp) had failure in balloon augmentation.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: b4, g3, g6, h2, h11.Corrected fields: b5, h6 (investigation finding).
|
| |
|
Event or Problem Description
|
|
It was reported that linear 7.5fr.34cc intra aortic balloon pump (iabp) had failure in balloon augmentation.There was no harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: b4, d9, g3, g6, h2, h3, h6 (type of investigation, investigation findings and investigation conclusions), h11.No decon or visual inspection performed since product was not returned and could not be evaluated.Therefore, we were unable to confirm or determine a root cause for the reported failure.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Search Alerts/Recalls
|