The device was inspected on-site by a dräger service engineer who could confirm the reported behavior and trace it back to a faulty motor of the ventilator.The entire motor assembly has been replaced; the device passed all consecutive tests and has been returned to use.Log file evaluation made by the manufacturer confirms the validity of the on-site expertise.The supervisor function of the software has detected speed deviations during motor rotation and has forced a shut-down of automatic ventilation.As confirmed for the particular case, the shut-down is accompanied by a corresponding alarm, and manual ventilation via the built-in breathing bag including dosage of all necessary gases including anesthetics is further possible.The safety concept can be explained as follows: the ventilator motor is a step motor that drives a piston i.E.The number of rotations and the end position defines the piston hub; the piston hub is an equivalent to the tidal volume applied to the patient.To protect the patient from potentially hazardous output and/or from serious damages to the ventilator unit, the device is designed to shut down automatic ventilation when the divergence between expected and measured motor position exceeds a certain limit.An in-depth investigation of the motor was not considered necessary since the issue could clearly be identified from the log entries.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|