| Catalog Number |
5536B400 |
| Medical Device Problem Codes |
Malposition of Device (2616); Migration (4003)
|
| Health Effect - Clinical Codes |
Inadequate Osseointegration (2646); Insufficient Information (4580)
|
| Date of Event |
12/31/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The following information was provided: left knee.Patient revised due to tibia component subsiding anteriorly.Surgery sheets and x-ray attached.No other information available due to hospital policy.
|
| |
|
Event or Problem Description
|
|
No new information.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding malposition involving a triathlon baseplate was reported.The event was confirmed via clinician review - loosening and wear were also confirmed through this review.Method & results
-product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: a review of the provided medical information by a clinical consultant indicated:
"the x-rays demonstrate a left tka with significant findings including marked differential in joint space between the medial and lateral compartments with near obliteration of the medial joint space.In addition, there is evidence of tibial subsidence consistent with loss of fixation.Mechanical failure of a total knee replacement secondary to wear and loss of tibial fixation is confirmed.The implant sheet from early 2025 confirms a revision surgery was completed but no other information regarding this surgery was provided.The root cause of these mechanical failures cannot be established from the information provided."
-product history review: review of the device history records could not be performed as lot details were not provided.-complaint history review:there have been no other similar events for the reported lot.Conclusions:
it was reported that 'patient revised due to tibia component subsiding anteriorly.'
a review of the provided medical information by a clinical consultant indicated:
"the x-rays demonstrate a left tka with significant findings including marked differential in joint space between the medial and lateral compartments with near obliteration of the medial joint space.In addition, there is evidence of tibial subsidence consistent with loss of fixation.Mechanical failure of a total knee replacement secondary to wear and loss of tibial fixation is confirmed.The implant sheet from early 2025 confirms a revision surgery was completed but no other information regarding this surgery was provided.The root cause of these mechanical failures cannot be established from the information provided."
no further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|