Model Number 1109598
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Device Sensing Problem (2917); Pressure Problem (3012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/05/2024
Type of Reportable Event Malfunction
Event or Problem Description
A bipap a40 ventilator was returned to a service center for remediation.There was no serious patient harm or injury reported.The device was not reported to be in clinical use.During the evaluation of the device at a service center, the device was visually inspected and found no evidence of foam degradation however, a ventilator inoperative error code indicating (the difference between the blower pressure sensor reading and the outlet pressure sensor reading is 10 cmh2o or greater for 5) was confirmed in the event log.No repair was performed.The device was scrapped as per customer request.This issue has been identified under the fsn 2023-cc-src-039 due to interruptions and/or loss of therapy due to a ventilation inoperative alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A40
Common Device Name
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key24159531
Report Number2518422-2026-003178
Device Sequence Number13043727
Product Code MNT
UDI-Device Identifier00606959042766
UDI-Public(01)00606959042766
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K121623
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1109598
Device Catalogue Number1109598
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2024
Initial Report FDA Received Date01/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/20/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-