| Model Number |
DDU-A2300 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported that their aed's asi is flashing red, and reporting a service code.The customer did not report this occurring during patient use.
|
| |
|
Additional Manufacturer Narrative
|
|
Although the customer initially reported a service code, review of the aed's internal log files determined that the service code after battery replacement was due to the previous battery fully depleting within the unit.Analysis of the aed's log files identified that the customer's failure to promptly address red asi warnings reduced battery life expectancy.As a follow up with the customer, the proper maintenance of this device was reviewed.Once the new battery was installed and a manual self test performed the aed functioned as designed and was able stay in service.
|
| |
|
Search Alerts/Recalls
|