| Model Number |
DDU-100A |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis of the aed's log files identified that the customer's failure to promptly address red asi warnings reduced battery life expectancy.As a follow up with the customer, the proper maintenance of this device was reviewed.Once the new battery was installed and a manual self test performed the aed functioned as designed and was able stay in service.
|
| |
|
Event or Problem Description
|
|
A customer reported that their aed's battery was depleted.When tested with a spare battery pack, it powered on and prompted a service code.The customer did not report this occurring during patient use.
|
| |
|