ABBOTT MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL0¿, 63 CM LENGTH, 8F; INTRODUCER, CATHETER
|
Back to Search Results |
|
| Model Number |
407449 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During the atrial fibrillation procedure, after insertion, air removal was attempted using a syringe; however, air continued to be aspirated.The introducer was replaced, and the procedure was completed with no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information d9, g3, g6, h2, h3, h6.One 8f swartz braided introducer sheath was received for evaluation.Functional testing determined a leak was noted in the hemostasis valve.The cap was removed from the hemostasis hub and the hemostasis seals were microscopically inspected.Tearing, resulting in a hole, was noted in the proximal and distal seals.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.
|
| |
|