• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY EVO ITALY; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY EVO ITALY; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number IT2110X21B
Medical Device Problem Code Pressure Problem (3012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2026
Type of Reportable Event Malfunction
Event or Problem Description
A ventilator was returned to a third-party service center for service with a reported issue of maintenance required.There was no serious patient harm or injury that occurred or was reported.During the evaluation of the device at the third-party service center, an error code related to pressure was observed in the error log.Evt_press_sensor_mismatch means the device software has detected a large pressure drop between the monitor and outlet pressure sensors (i.E.Across the machine flow sensor).The device evaluation does not identify any specific component malfunction that would need to be replaced to resolve the issue, but necessary parts were replaced as if preventive maintenance was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO ITALY
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key24162703
Report Number2518422-2026-003191
Device Sequence Number9368902
Product Code CBK
UDI-Device Identifier00606959055421
UDI-Public(01)00606959055421
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberIT2110X21B
Device Catalogue NumberIT2110X21B
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/22/2026
Initial Report FDA Received Date01/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/09/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-