• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP MIS 16CM CURVED ATTACHMENT; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP MIS 16CM CURVED ATTACHMENT; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Catalog Number 5407120970C
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between october 1 ¿ december 31 2025.This report summarizes 10 events for the failure: overheating of device.7 events had problem identified during non-clinical procedure; there was no patient involvement.3 events had no health consequences or impact.
 
Additional Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events 10 events were reported for this quarter.Product return status 10 devices were evaluated based on historical data analysis.Evaluation findings (results/components) 10 devices: degradation problem identified, wear problem, lubrication problem identified, overheating problem identified / part/component/sub-assembly term not applicable.Product disposition 3 devices: product not received.7 devices: scrapped by stryker.Additional information 10 devices were not labeled for single-use.10 devices were not reprocessed or reused.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIS 16CM CURVED ATTACHMENT
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
1941 stryker way
portage MI 49002
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
kelsey barla
1941 stryker way
portage, MI 49002
2693237700
MDR Report Key24163088
Report Number3015967359-2026-99030
Device Sequence Number9252168
Product Code HBE
UDI-Device Identifier07613327096521
UDI-Public07613327096521
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Exemption NumberVMSR
Number of Events Summarized10
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number5407120970C
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/31/2025
Initial Report FDA Received Date01/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-