STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP DUAL TRIGGER LARGE COLLET; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 8203226000
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 5 events were reported for this quarter.Product return status: 5 devices were evaluated in the field.Evaluation findings (results/components): 5 devices: material and/or chemical problem identified / coating material.Product disposition: 5 devices: product disposition is not yet determined.Additional information: 5 devices were not labeled for single-use.5 devices were not reprocessed or reused.
 
Event or Problem Description
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between october 1 ¿ december 31, 2025.This report summarizes 5 events for the failure: flaked.- 5 events had problem identified during non-clinical procedure; there was no patient involvement.
 
Additional Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Mfr report # 3015967359-2026-99186.Supplemental rationale: corrected data: h11, 5 previously reported events were included in this follow-up record.Product return status: 5 devices were evaluated in the field.Evaluation findings (results/components): 5 devices: material and/or chemical problem identified / coating material.Product disposition: 5 devices: scrapped by stryker.
 
Event or Problem Description
No change.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL TRIGGER LARGE COLLET
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
1941 stryker way
portage MI 49002
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
kelsey barla
1941 stryker way
portage, MI 49002
2693237700
MDR Report Key24164265
Report Number3015967359-2026-99186
Device Sequence Number9001223
Product Code HWE
UDI-Device Identifier07613327304442
UDI-Public07613327304442
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Exemption NumberVMSR
Number of Events Summarized5
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number8203226000
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/31/2025
Supplement Date Received by Manufacturer12/31/2025
Initial Report FDA Received Date01/23/2026
Supplement Report FDA Received Date01/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-