| Model Number |
CI-1601-05 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
02/17/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Event or Problem Description
|
|
The recipient is reportedly experiencing sound quality issues.Programming adjustments were made, however, the issue did not resolve.A ct confirmed the recipient is reportedly experiencing device migration.Revision surgery will be scheduled.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information section: b.3, d.6b, d.9, h.6.The recipient's device was explanted due to electrode migration.The recipient was re-implanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: section b1 & h1.Additional information: section h.6.Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed that the electrode was severed and silicone damage was observed on the top and bottom cover of the device.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The condition of the electrode prevented an electrical test from being performed.The device passes the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|