Catalog Number D128208
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/22/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to johnson & johnson medtech (j&j medtech) for evaluation 26-jan-2026.As such, section d9 has been populated.It was reported that during operation, the device was not irrigating.Device investigation details: following j&j medtech procedures, a visual inspection and irrigation test of the returned device were performed.Visual inspection revealed no damage or anomalies on the device.An irrigation test was performed, and the device was irrigating correctly, no obstructed holes were observed.No irrigation issues were observed.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The irrigation issue reported by the customer could not be replicated during the product investigation.Other issues or circumstances may have occurred during the use of the device that may have affected its performance.The instructions for use (ifu) contain the following information: flush the catheter with heparinized saline prior to insertion into the body.Always follow standard practices of using a continuous drip of anticoagulant fluid under pressure through the proximal luer connector when the device is in the body.As part of johnson & johnson medtech's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient underwent an atrial fibrillation (afib) ¿ paroxysmal ablation procedure with a pentaray nav.Initially, it was indicated that during operation, the device (including port, luer hub) was not irrigating.A second device was used to complete the operation.There was no adverse event reported on patient.With the initial information available, this event was assessed as non mdr reportable.However, on 28-dec-2025, additional information was received which confirmed that the suspected device is the catheter and that it was used in the patient.Therefore, the event was re-assessed as mdr reportable with an awareness date of 28-dec-2025.It was also indicated that the ¿head end does not produce water¿.A pressure belt was used (no pump was used).The steps taken to resolve the irrigation issue were pressurization and flushing.The correct catheter settings were selected on the generator.There was no issue with flow rate change at the start of ablation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTARAY NAV
Common Device Name
CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
9497882078
MDR Report Key24168660
Report Number2029046-2026-00308
Device Sequence Number18454611
Product Code MTD
UDI-Device Identifier10846835012224
UDI-Public(01)10846835012224(17)270910(10)31444618L
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K123837
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberD128208
Device Lot Number31444618L
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/28/2025
Supplement Date Received by Manufacturer01/26/2026
Initial Report FDA Received Date01/23/2026
Supplement Report FDA Received Date02/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-