• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO VILLALBA CAPSURE SENSE LEAD MRI SURESCAN; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO VILLALBA CAPSURE SENSE LEAD MRI SURESCAN; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 457453
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Thrombosis/Thrombus (4440); Medical device site infection (4845)
Date of Event 02/26/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Continuation of d10: 407658 lead implanted: (b)(6) 2009, 429878 lead implanted: (b)(6) 2019.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) system was explanted and replaced due to infection.Vegetation was suspected.The patient experienced deep vein thrombosis (dvt) and medication was required. no further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSURE SENSE LEAD MRI SURESCAN
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key24169059
Report Number2649622-2026-02224
Device Sequence Number18419762
Product Code DTB
UDI-Device Identifier00885074200958
UDI-Public(01)00885074200958(17)110119
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P830061
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 01/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date01/19/2011
Device Model Number457453
Device Catalogue Number457453
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/24/2025
Initial Report FDA Received Date01/23/2026
Date Device Manufactured01/27/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DTPA2QQ CRT-D, 6935M55 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age78 YR
Patient SexMale
Patient Weight75 KG
-
-