| Model Number |
420172-18 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was a broken cable with a da vinci product.The customer reported event has no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The maryland bipolar forceps instrument was analyzed and found to have a broken conductor wire within the bipolar yaw pulley, between the wire crimp on the grip and the silicone potting, which seals the wire entrance to the yaw pulley.The instrument failed the electrical continuity test.No thermal damage was found on the instrument.Components adjacent to the broken wire do not show damage.Common causes of broken instrument conductor wire - bipolar are attributed to damage through external collisions, during use, or during reprocessing.
|
| |
|