| Model Number |
UC8519 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a clearlink continu-flo solution set with duo-vent spike was leaking from the connection the total parenteral nutrition (tpn) filter tubing and the triple port.It was unknown if a patient was connected during this event.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to h6.H11: the actual device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Pressure testing was performed, and it was noted that there was a leak on the edge of the blue part of the chamber.The reported condition was verified.The cause of the reported condition was a manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|