• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PARKER MEDICAL PARKER; ENDOTRACHEAL TUBE, HIGH VOLUME LOW PRESSURE (PFHV), CUFFED, PARKER, 8.0MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PARKER MEDICAL PARKER; ENDOTRACHEAL TUBE, HIGH VOLUME LOW PRESSURE (PFHV), CUFFED, PARKER, 8.0MM Back to Search Results
Model Number H-PFHV-80 (I-PFHV-80)
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product involved in the report has not been returned.A review of the device history record is in-progress all information reasonably known as of 23jan 2026 has been included in the health authority report.Should additional information be obtained a follow-up health authority report will be provided.The information provided by airlife.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the airlife complaint database and is identified as complaint -(b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a airlife product is defective or causes serious injury.
 
Event or Problem Description
It was reported:that during intubation, the inflating tube detached. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARKER
Common Device Name
ENDOTRACHEAL TUBE, HIGH VOLUME LOW PRESSURE (PFHV), CUFFED, PARKER, 8.0MM
Manufacturer (Section D)
PARKER MEDICAL
9457 s university blvd
pmb 331
highlands ranch CO 80126
Manufacturer (Section G)
WELL LEAD MEDICAL CO
c-4 jinhu industrial estate
hualong
panyu, guangzhou 51143 4
JA   511434
Manufacturer Contact
melissa brickley
2710 northridge dr nw suite a
grand rapids, MI 49544
6162598415
MDR Report Key24172859
Report Number3004748541-2026-00017
Device Sequence Number9557694
Product Code BTR
UDI-Device Identifier00607411952753
UDI-Public(01)00607411952753
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K984528
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH-PFHV-80 (I-PFHV-80)
Device Catalogue NumberH-PFHV-80 (I-PFHV-80)
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/25/2025
Supplement Date Received by Manufacturer12/25/2025
Initial Report FDA Received Date01/23/2026
Supplement Report FDA Received Date05/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-