| Model Number |
H74939135204010 |
| Medical Device Problem Codes |
Material Rupture (1546); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in the moderately tortuous and severely calcified vessel below the knee.A 2mm x 40mm x 145cm coyote es balloon was advanced for dilation.However, during the first inflation at 6 atmospheres for 15 seconds, the balloon ruptured.The device was removed and the procedure was completed with a different device.There were no patient complications as a result of this event, and the patient was in good condition post-procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in the moderately tortuous and severely calcified vessel below the knee.A 2mm x 40mm x 145cm coyote es balloon was advanced for dilation.However, during the first inflation at 6 atmospheres for 15 seconds, the balloon ruptured.The device was removed and the procedure was completed with a different device.There were no patient complications as a result of this event, and the patient was in good condition post-procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Device analysis findings: upon receipt at our post market quality assurance laboratory, the outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple buckling along the shaft.There is a longitudinal tear in the balloon 7mm from the tip and is approximately 20mm long.Microscopic examination revealed no additional damages.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed there is a longitudinal tear in the balloon.Device history record: it was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of adverse event related to patient condition.
|
| |
|