| Model Number |
H74939135304010 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1 - initial reporter address 1: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the moderately tortuous and moderately calcified vessel below the knee.A 3mm x 40mm x 146cm coyote es balloon was advanced for dilation.However, during the second inflation at 14 atmospheres for 60 seconds, the balloon ruptured.The device was removed and the procedure was completed with a different device.There were no patient complications as a result of this event, and the patient was in good condition post-procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the moderately tortuous and moderately calcified vessel below the knee.A 3mm x 40mm x 146cm coyote es balloon was advanced for dilation.However, during the second inflation at 14 atmospheres for 60 seconds, the balloon ruptured.The device was removed and the procedure was completed with a different device.There were no patient complications as a result of this event, and the patient was in good condition post-procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
E1 - initial reporter address 1: (b)(6).Device analysis findings: upon receipt at our post market quality assurance laboratory, the outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a pinhole 22mm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed there is a pinhole in the balloon.Device history record: it was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of adverse event related to patient condition.
|
| |
|
Search Alerts/Recalls
|