|
It was reported, during a transurethral ureteroscopy, lithotripsy, and stone removal procedure, a basket wire from an ncircle delta wire tipless stone extractor broke inside the flexible sheath, rendering the basket unusable.The device was tested prior to use and functioned properly.The device was successfully used approximately five times to retrieve stones before the issue occurred.A new basket was used to complete the procedure.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
|
|
E1 - phone number: (b)(6).G4 ¿ pma/510(k) #: exempt.H3 - device has been returned and preliminary evaluation has been performed, however, our investigation is ongoing and device evaluation summary will be included in our follow up report once our investigation has been completed.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation ¿ evaluation.It was reported, during a transurethral ureteroscopy, lithotripsy, and stone removal procedure, a basket wire from an ncircle delta wire tipless stone extractor broke inside the flexible sheath, rendering the basket unusable.The device was tested prior to use and functioned properly.The device was successfully used approximately five times to retrieve stones before the issue occurred.A new basket was used to complete the procedure.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturer¿s instructions (mi), and quality control (qc) procedures were conducted during the investigation.Visual inspection of the returned complaint device was also conducted.One device was returned for investigation in an open package with label.One wire, in the basket formation, was broken.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record showed no discrepancies related to the reported failure mode.A complaint history search identified no other events reported for the related failure mode.According to the available information, cook has concluded that the device was manufactured to specification and that there is no evidence suggesting nonconforming product exists either in house or in the field.Cook also reviewed product labeling.The ifu supplied with the device did not provide any information related to the reported issue.Based upon the available information, inspection of the returned device, and results of the investigation, cook has concluded that the cause for the complaint could not be established.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints per the risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|