| Catalog Number |
08469865160 |
| Medical Device Problem Codes |
Low Test Results (2458); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The serial number of the e 601 analyzer is (b)(6).The field service engineer cleaned the inside of the sipper probes and pre-wash system tubing.Probes were externally cleaned.Maintenance actions, including a probe washes, mixer rinsing, pre-wash system rinsing, reagent priming, and air purges were performed.Measuring cell preparation maintenance was performed.A system volume check, photomultiplier tube high voltage adjustment, and blank cell calibration were performed.Performance testing was run.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one patient sample tested with elecsys troponin t gen 5 stat on a cobas 6000 e601 module.The sample initially resulted in a troponin t value of 11.12 ng/l.The patient's nurse asked for a repeat of the sample since the value did not agree with the patient's clinical status.The sample was repeated, resulting in a value of 115.2 ng/l with a data flag.The repeat value was deemed correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation included a review of the data set provided by the customer: the qc results were within the specified ranges.There was no indication of a performance issue with the reagent.The alarm trace contained "abnormal probe sucking" and "sample short" errors.These are indicators of possible poor sample quality.A general reagent problem was not detected because the qc before the event was within ranges, and isolated non-reproducible results do not represent a general malfunction of the assay.The investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|
Search Alerts/Recalls
|