• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH TFNA FENESTRATED SCREW 95MM - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH TFNA FENESTRATED SCREW 95MM - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 04.038.195S
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, while conducting a tfna, surgeon attempted to place a 95mm screw over our 3.2mm guide wire an it would not pass.Surgeon took the screw off the insertion handle, passed the 3.2 guide wire independently through the screw, screw inserted and retention pin without issue.Surgeon then reassembled the screw onto the screw inserter with the connecting screw and it again did not pass.Surgeon opened another tfna locking set and the same issue occurred.Surgeon then opened a new 95mm tfna screw and it worked without any issue.There was no patient impact for this issue.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.Investigation summary: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot: manufacturing location: elmira ¿ packaged, labeled, sterilized and released by: monument.Manufacturing date: 26-jun-2025.Expiration date: 31-may-2035.Part number: 04.038.195s, tfna fenestrated screw 95mm -sterile.Lot number: 71337p8 (sterile).Lot quantity: (b)(4).Nonconformance noted: n/a.Review of the dhr file: component part manufactured by elmira; lot number 70101p2 quantity (b)(4).Production order traveler met all inspection acceptance criteria.Packaging label log (pll), lppf rev g, lmd rev b was reviewed and was determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Scn (b)(4) supplied by (b)(4) was reviewed and determined to be conforming.A manufacturing record evaluation was performed for the finished device with batch number 71337p8.No nonconformities or manufacturing irregularities have identified related to the complaint condition.Device history batch: null.Device history review: a manufacturing record evaluation was performed for the finished device with batch number 71337p8.No nonconformities or manufacturing irregularities have identified related to the complaint condition.If additional information is made available, the investigation will be updated as applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TFNA FENESTRATED SCREW 95MM - STERILE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
elba bello
1302 wrights lane east
west chester, PA 19380
9083863534
MDR Report Key24177864
Report Number8030965-2026-00858
Device Sequence Number18636801
Product Code HSB
UDI-Device Identifier10886982098757
UDI-Public(01)10886982098757(17)350531(10)71337P8
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K160167
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.038.195S
Device Lot Number71337P8
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/12/2026
Supplement Date Received by Manufacturer02/19/2026
Initial Report FDA Received Date01/26/2026
Supplement Report FDA Received Date02/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/26/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-