| Catalog Number |
6506000000 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between october 1-december 31, 2025.This report summarizes 5 malfunction event(s), where it was reported the device experienced inability to disengage the cot from the cot fastener.No adverse consequence or clinically relevant delay in treatment was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as part of the fda voluntary.Malfunction summary reporting program.5 device(s) was/were functionally/visually.Inspected in the field.The device(s) was/were repaired and returned to use.Evaluation.Results findings (components/results) 1 device: chassis/frame; fastener / deformation.Problem: mechanical problem identified 1 device: chassis/frame / deformation problem 1.Device: controller / wear problem 2 devices: chassis/frame / mechanical problem identified.There was no remedial action taken.This device is not labeled for single use.
|
| |
|