• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. 2008T HEMODIALYSIS SYSTEM W/BIBAG; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. 2008T HEMODIALYSIS SYSTEM W/BIBAG; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 190766
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event or Problem Description
A user facility's biomedical technician (bmt) reported that a 2008t hemodialysis (hd) machine displayed an a.11 alarm during set up.Upon follow up, the bmt said that he replaced the lp 956 and lp 955 boards to resolve the reported issue.The machine is back in service.He also said that the lp956 board showed signs of thermal decomposition.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.The complaint device is not available to be returned to the manufacturer for evaluation.Multiple attempts have been made to contact the customer to obtain additional information related to the reported event.At this time, no additional information has been provided.
 
Event or Problem Description
A user facility's biomedical technician (bmt) reported that a 2008t hemodialysis (hd) machine displayed an a.11 alarm during set up.Upon follow up, the bmt said that he replaced the lp 956 and lp 955 boards to resolve the reported issue.The machine is back in service.He also said that the lp956 board showed signs of thermal decomposition.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.The complaint device is not available to be returned to the manufacturer for evaluation.Multiple attempts have been made to contact the customer to obtain additional information related to the reported event.At this time, no additional information has been provided.
 
Additional Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation and an on-site evaluation was not performed by a fresenius field service technician (fst).The manufacturer was able to determine a causal relationship between the objective evidence provided by the customer and the reported event.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The reported event has been confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HEMODIALYSIS SYSTEM W/BIBAG
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key24178457
Report Number0008030665-2026-00175
Device Sequence Number9578601
Product Code KDI
UDI-Device Identifier00840861100910
UDI-Public00840861100910
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K150708
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 03/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number190766
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/22/2026
Supplement Date Received by Manufacturer03/09/2026
Initial Report FDA Received Date01/26/2026
Supplement Report FDA Received Date03/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/20/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-