|
|
| Catalog Number |
311.43 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: b5, d10 (concomitant) corrected: h8 if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received: a.Date of event (if available): - items identified at various times in preop testing over the last few months.B.What is the timing of the reported event? (pre-op, intra-op, post-op) - pre-op.
|
| |
|
Event or Problem Description
|
|
It was reported multiple devices are no longer functional.These issues did not impact patient outcomes.Below devices were found to be "release mechanism broken" during inspection: 03.043.028.Below devices were found to be "stripped threads" during inspection: 03.120.022, 312.648, 323.029.Below devices were found to be "nut stuck on sleeve" during inspection: 03.037.017, 03.037.016.Below devices were found to be "handle leaching/spd wil not sterilize" during inspection: 03.100.018.Below devices were found to be "broken handle" during inspection: 311.43.Below devices were found to be "stripped drill connection and 2nd is cold welded to a schanz pin" during inspection: 393.103.Below devices were found to be "broken sliding mechanism" during inspection: 319.091.Below devices were found to be "not snapping in" during inspection: 03.120.018.03.120.017 below devices were found to be "stripped" during inspection: 03.010.151, 314.467, 03.900.042, 03.130.010, 314.453, 03.010.150, 03.010.152.Below devices were found to be "not working properly" during inspection: 329.15.Below devices were found to be "not locking tension between trigger pulls" during inspection: 03.221.015.Below devices were found to be "not aligned" during inspection: 398.65.Below devices were found to be "not locking into aiming arm" during inspection: 03.231.007.Below devices were found to be "bent handle" during inspection: 393.1.Below devices were found to be "trigger detached from handle" during inspection: 03.221.007.Below devices were found to be "not sliding properly" during inspection: 319.006.Below devices were found to be "shaft pulls out of handle" during inspection: 314.118.Below devices were found to be "not locking on" during inspection: 351.16j.Below devices were found to be "binding on qc attachments" during inspection: 511.773.Below devices were found to be "ends not aligning" during inspection: 03.221.011.(b)(4) is the mother complaint.(b)(4) are related to each other and are created to capture additional ips.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d10.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d9.H3, h6: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found the brown handle was broken from the distal area.Broken fragments were not returned.Lot number was not observed over the surface nor handle of the device, this most likely to manufacturing date.The observed condition of the device was consistent with a component failure that was caused by exposure to unintended forces.Properly handling and attention to the approved use of the device diminishes the risk of failure.A functional test was unable to be performed due to the post-manufacturing damage.A dimensional inspection was unable to be performed due to the post-manufacturing damage.The overall complaint was confirmed as the observed condition of the handle with quick coupling, length 110 mm would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the device lot number is unknown, therefore a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|