| Model Number |
310-2421312-01 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device was unable to detect the attached electrode pads.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation; the customer's report was observed and attributed to the multifunction cable (mfc) harness which was observed disconnected from a connector of the monitor board.The mfc harness was reseated to correct the reported problem.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|