MAUDE Adverse Event Report: SYNTHES GMBH DEPTH GAUGE F/CORTSCR Ø3.5 MEAS-RANGE UP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH DEPTH GAUGE F/CORTSCR Ø3.5 MEAS-RANGE UP Back to Search Results
Catalog Number 319.091
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported multiple devices are no longer functional.These issues did not impact patient outcomes.Below devices were found to be "release mechanism broken" during inspection: 03.043.028.Below devices were found to be "stripped threads" during inspection: 03.120.022, 312.648, 323.029.Below devices were found to be "nut stuck on sleeve" during inspection: 03.037.017, 03.037.016.Below devices were found to be "handle leaching/spd wil not sterilize" during inspection: 03.100.018.Below devices were found to be "broken handle" during inspection: 311.43.Below devices were found to be "stripped drill connection and 2nd is cold welded to a schanz pin" during inspection: 393.103.Below devices were found to be "broken sliding mechanism" during inspection: 319.091.Below devices were found to be "not snapping in" during inspection: 03.120.018, 03.120.017.Below devices were found to be "stripped" during inspection: 03.010.151, 314.467, 03.900.042, 03.130.010, 314.453, 03.010.150, 03.010.152.Below devices were found to be "not working properly" during inspection: 329.15.Below devices were found to be "not locking tension between trigger pulls" during inspection: 03.221.015.Below devices were found to be "not aligned" during inspection: 398.65.Below devices were found to be "not locking into aiming arm" during inspection: 03.231.007.Below devices were found to be "bent handle" during inspection: 393.1.Below devices were found to be "trigger detached from handle" during inspection: 03.221.007.Below devices were found to be "not sliding properly" during inspection: 319.006.Below devices were found to be "shaft pulls out of handle" during inspection: 314.118.Below devices were found to be "not locking on" during inspection: 351.16j.Below devices were found to be "binding on qc attachments" during inspection: 511.773.Below devices were found to be "ends not aligning" during inspection: 03.221.011.(b)(4) is the mother complaint.(b)(4) are related to each other and are created to capture additional ips.
 
Additional Manufacturer Narrative
Prodproduct complaint #(b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d4 (lot, expiration date), d9, h4 corrected: d4 primary udi number the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: b5, d10 (concomitant) if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Additional information received: a.Date of event (if available): items identified at various times in preop testing over the last few months.B.What is the timing of the reported event? (pre-op, intra-op, post-op) pre-op.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h3, h6: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found only the slider and the depth gauge body with hook were returned.Breakage was not identified, however, due to the damage observed, the reported event can be confirmed.The observed condition of the device can be consistent with a component failure that was caused by exposure to unintended forces.However, with available evidence, the complete potential cause can not be determined.A functional test was unable to be performed due to the post-manufacturing damage.A dimensional inspection was unable to be performed due to the post-manufacturing damage.The overall complaint was confirmed as the observed condition of the depth gauge f/cortscr ø3.5 meas-range up would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: part: 319.091, synthes lot: j000666, supplier lot: j000666, supplier: (b)(4), release to warehouse date: aug 02, 2021.No ncrs generated during production.Device history batch: null.Device history review: review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d10.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE F/CORTSCR Ø3.5 MEAS-RANGE UP
Common Device Name
DEPTH GAUGE
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key24179477
Report Number8030965-2026-00874
Device Sequence Number9524217
Product Code HTJ
UDI-Device Identifier10886982190000
UDI-Public(01)10886982190000(17)411231(10)J000666
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number319.091
Device Lot NumberJ000666
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/12/2026
Supplement Date Received by Manufacturer01/28/2026
02/24/2026
02/27/2026
03/16/2026
Initial Report FDA Received Date01/26/2026
Supplement Report FDA Received Date02/02/2026
02/26/2026
03/02/2026
03/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BALL SPIK STRAIG W/POINTED BALL TIPS Ø6.; CABLE TENSIONER ONE-H+OPERABLE.; CERCL-PASSER Ø60 MIN-INVASIVE.; FLEX SFT CONNECTOR FOR USE JACOBS CHUCK.; FORCEPS FOR BROKEN SCREW REMOVAL.; GUIDE SLEEVE F/AIMING ARM INSTR F/VA-LCP.; HANDLE WITH QUICK COUPLING, LENGTH 110 MM.; TORQUE LIMITER 1.5NM F/AO/ASIF.; UNIV-CHUCK W/T-HANDLE.