• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH BALL SPIK STRAIG W/POINTED BALL TIPS Ø6.; RETRACTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH BALL SPIK STRAIG W/POINTED BALL TIPS Ø6.; RETRACTOR Back to Search Results
Catalog Number 03.100.018
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported multiple devices are no longer functional.These issues did not impact patient outcomes.Below devices were found to be "release mechanism broken" during inspection: 03.043.028.Below devices were found to be "stripped threads" during inspection: 03.120.022, 312.648, 323.029.Below devices were found to be "nut stuck on sleeve" during inspection: 03.037.017, 03.037.016.Below devices were found to be "handle leaching/spd wil not sterilize" during inspection: 03.100.018.Below devices were found to be "broken handle" during inspection: 311.43.Below devices were found to be "stripped drill connection and 2nd is cold welded to a schanz pin" during inspection: 393.103.Below devices were found to be "broken sliding mechanism" during inspection: 319.091.Below devices were found to be "not snapping in" during inspection: 03.120.018, 03.120.017.Below devices were found to be "stripped" during inspection: 03.010.151, 314.467, 03.900.042, 03.130.010, 314.453, 03.010.150, 03.010.152.Below devices were found to be "not working properly" during inspection: 329.15.Below devices were found to be "not locking tension between trigger pulls" during inspection: 03.221.015.Below devices were found to be "not aligned" during inspection: 398.65.Below devices were found to be "not locking into aiming arm" during inspection: 03.231.007.Below devices were found to be "bent handle" during inspection: 393.1.Below devices were found to be "trigger detached from handle" during inspection: 03.221.007.Below devices were found to be "not sliding properly" during inspection: 319.006.Below devices were found to be "shaft pulls out of handle" during inspection: 314.118.Below devices were found to be "not locking on" during inspection: 351.16j.Below devices were found to be "binding on qc attachments" during inspection: 511.773.Below devices were found to be "ends not aligning" during inspection: 03.221.011.(b)(4) is the mother complaint.(b)(4) are related to each other and are created to capture additional ips.
 
Additional Manufacturer Narrative
Pro dproduct complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d4 lot, d9.Corrected: d4 primary udi number, g1.H3, h6: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found foreign substance was not identified.However the handle was observed with orange spots near the two pin indicating a possible corrosion.The reported condition can be confirmed.Potential cause can be traced to maintenance, refer to the device/country specific ifu for information related to cleaning, sterilization, reprocessing instructions, and inspection procedures.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the ball spik straig w/pointed ball tips ø6.Would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: part number: 03.100.025, lot number: a70a25, manufacturing site: tuttlingen, release to warehouse date: week 25, 2005.A review of the device history records is not possible, the dhr is no longer available due to the age of the instrument.Information retrieved from (b)(4) as the impacted products share the same lot number.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: b5, d10 (concomitant).Corrected: h8.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Additional information received: a.Date of event (if available): - items identified at various times in preop testing over the last few months.B.What is the timing of the reported event? (pre-op, intra-op, post-op) - pre-op.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d10.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALL SPIK STRAIG W/POINTED BALL TIPS Ø6.
Common Device Name
RETRACTOR
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
SYNTHES TUTTLINGEN
unter hasslen 5
tuttlingen CO
GM  
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key24179635
Report Number8030965-2026-00875
Device Sequence Number18531118
Product Code GAD
UDI-Device Identifier10886982070678
UDI-Public(01)10886982070678(10)A70A25
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.100.018
Device Lot NumberA70A25
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/12/2026
Supplement Date Received by Manufacturer01/28/2026
02/24/2026
03/16/2026
Initial Report FDA Received Date01/26/2026
Supplement Report FDA Received Date02/02/2026
02/26/2026
03/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CABLE TENSIONER ONE-H+OPERABLE.; CERCL-PASSER Ø60 MIN-INVASIVE.; DEPTH GAUGE F/CORTSCR Ø3.5 MEAS-RANGE UP.; FLEX SFT CONNECTOR FOR USE JACOBS CHUCK.; FORCEPS FOR BROKEN SCREW REMOVAL.; GUIDE SLEEVE F/AIMING ARM INSTR F/VA-LCP.; HANDLE WITH QUICK COUPLING, LENGTH 110 MM.; TORQUE LIMITER 1.5NM F/AO/ASIF.; UNIV-CHUCK W/T-HANDLE.
Patient SexUnknown
-
-